Technical data
16.5
Fresh-gas delivery with electronically controlled gas
mixer (continued)
Information
Accuracy
16.6
Fresh-gas delivery with mechanically controlled gas
mixer
Standardization
O
concentration
2
Setting range for fresh-gas flow
Electronic measurement of fresh-gas flow
Range
Accuracy
Resolution of the value displayed on the screen
Resolution of the value displayed in the status
display
Total flow tube
Standardization
Range
Accuracy
Resolution of displayed value
16.7

Ventilator

Design
Time-based settings
Respiratory rate RR
270
The maximum flow specified under STPD con-
ditions can drop linearly from 15 L/min at
750 hPa ambient pressure to 13.5 L/min at
620 hPa ambient pressure. At 620 hPa ambient
pressure, an STPD flow of 13.5 L/min corre-
sponds to an ATPD flow of 22 L/min.
±10 % or ±50 mL/min (the larger value applies)
All data are standardized to STPD conditions.
21 to 100 Vol%
0 to at least 12 L/min (O
0 to 15 L/min (O
±10 % or ±0.12 L/min (the larger value applies)
0.01 L/min (from 0 to 0.2 L/min)
0.02 L/min (from 0.2 to 0.5 L/min)
0.05 L/min (from 0.5 to 1 L/min)
0.10 L/min (from 1 to 15 L/min)
0.1 L/min
All data are standardized to STPD conditions.
0 to 10 L/min
±10 % of the set value at 100 % O
flows >1 L/min
0.5 L/min
Electronically driven ventilator, fresh-gas
decoupled
3 to 100 /min
±10 % of the set value or ±1 /min (the larger
value applies)
|
Instructions for use
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
, Air, and N
O)
2
2
, Air, and N
O)
2
2
and for
2