Functional Scope - Dräger Atlan Quick Reference Manual

Table of Contents

Advertisement

Overview
1.2

Functional scope

1.2.1
Safety functions
1.2.1.1
Overview
The following table gives an overview of the integrated safety functions which come into
effect if problems occur during operation:
Fault
Leakage
Mains power supply
failure
Mains power supply
failure and battery dis-
charged
Failure of the central
gas supply system
Complete failure of
the gas supply
Flow measurement
failure
Failure of fresh-gas
delivery (electroni-
cally controlled mixer
only)
Ventilator failure
Screen fault (screen
does not respond to
operation or has
failed)
14
Safety function
Pressure regulation and PEEP compensation
In mechanical ventilation modes, the set PEEP is actively
adjusted even in the case of small leaks.
In pressure-controlled modes, the pressure is controlled
during the entire respiratory cycle.
Uninterruptible power supply provided by internal battery
Device runtime in battery operation see Instruction for Use
Automatic deactivation of the breathing system warmer to
increase battery runtime
Manual ventilation and spontaneous breathing are available
Emergency O
delivery (electronically controlled gas mixer)
2
Fresh-gas delivery (mechanically controlled gas mixer)
Delivery of anesthetic agents via connected vaporizers
Use of the connected gas cylinders
Mechanical ventilation with ambient air possible (the hose
with the breathing bag will have to be removed for this)
No anesthetic agent delivery from the connected vaporizers
possible; switch to intravenous anesthetic agent required
Mechanical ventilation can be continued.
Restrictions with regard to displayed measured values, mea-
surement accuracies, and when triggering mandatory
breaths are possible.
Emergency O
delivery
2
see Instruction for Use
Anesthetic agent delivery from connected vaporizers possi-
ble
All ventilation modes are available.
Alternatively:
Backup manual mode
Manual ventilation or spontaneous breathing possible
Fresh-gas delivery available
Anesthetic agent delivery from connected vaporizers possi-
ble
Alternatively:
Backup manual mode
If backup manual mode is activated:
Manual ventilation or spontaneous breathing possible
Emergency O
delivery (electronically controlled gas mixer)
2
or fresh-gas delivery (mechanically controlled gas mixer)
available
Anesthetic agent delivery from connected vaporizers possi-
ble
Quick reference guide
|
Atlan SW 2.0n