Biomedical Support

This section of the appendix is designed for the Biomedical personnel of your
hospital. It includes an overview of the Biomed menu, basic setup, saving setups,
diagnostic functions, and calibration check procedures for NBP. This appendix is not a
substitute for the Infinity Vista Service Manual that is available from Dräger.
CAUTION: Verify that the monitor's line frequency under the "Service" menu is set to
match the 50 or 60 Hz mains power line frequency of your hospital. The line frequency
determines the center frequency of the ECG notch filter. The filter greatly reduces line
frequency noise on the ECG waveform. Refer to the Service manual for details how to
change frequency.

Startup Tests

The monitor runs internal tests continuously and upon power up to check various
functions, as well as the integrity of both ROM and RAM memories. If any of the tests
fail, the monitor resets and displays a message (see the following table). The test
results are recorded in the Diagnostic Log.
If the same error is detected three times in succession within 10 minutes without any
other intervening errors, the monitor stops resetting and generates a loud, continuous
error tone. If this happens, take the unit out of operation and call DrägerService.
NOTE: The monitor stores patient data and settings in an internal battery backed-up
memory. If any of the following messages appear frequently, the internal battery needs
to be changed.
Message
Default Patient Settings Restored
Software Reset, See Diagnostic
Log
Replace Battery Pack
Check Internal Battery
Calendar Clock Reset
Low Battery Reset
High Temperature Reset
NBP Fault
Patient Data Erased
C-8
B
S
IOMEDICAL
UPPORT
Test
Integrity of patient data settings.
The software has previously detected and logged a fault
condition. See the Diagnostic log for the exact cause of the
reset.
Battery <25%.
Assure integrity of patient data and/or settings.
Invalid clock setting at startup.
Voltage < 10V.
Internal temperature > 64°C.
Integrity of the NBP pneumatic constants.
Integrity of patient data.