Criteria
If the DA Module is not operating properly, change the DA module.
3.8

Lens Calibration

NOTE: After repairing the unit or changing the Engine module, the Lens calibration is required.

Procedure

1. Put the projector on a horizontal surface.
2. Press M
ENU
3. Select Configuration > Lens Setting.
4. Select Lens Calibration.

Inspection item

Check if the Lens Shift Module is moving smoothly.
Criteria
If the Lens Shift Module is not moving smoothly or making noises, please change the Engine module.
3.9

ADC Calibration

If the color is abnormal after upgrading the firmware, execute ADC calibration.

PC Calibration Procedure

1. Test signal: 1920 x 1080@60Hz
2. Test Pattern: Grays 16
3. Calibration pattern should be in full screen mode, navigate to:
Main Menu > Size & Position > Size Presets > Full Size.
4. In factory mode, enter ADC Calibration, and press the H
calibration.
Inspection item
Check if there are lines or noise on the screen.

Video Calibration Procedure

1. Test signal: 480i
2. Test Pattern: SMPTE BAR
3. Calibration pattern should be in full screen mode, navigate to:
Main Menu > Size & Position > Size Presets > Full Size.
4. In factory mode, enter ADC Calibration, and press the H
calibration.
E Series 675/775 Service Manual
020-000546-04 Rev. 1 (04-2017)
key to enter the Main menu.
Section 3: Service Setup
ELP
key to begin
Grays 16
ELP
key to begin
3-9