System Subsystem - R&S SMA100A Application Notes

Remote emulation modes with signal generators
Table of Contents

Advertisement

1GP71_0E
HP8360 emulation mode
Status
Command / Function / Remark / Limitation
SWEep:STEP value
Set / Query the size of each frequency step
Value:
num [HZ | KHZ | MHZ | GHZ] | MAXimum | MINimum
0.0 Hz _ num _ Fmax
Default:
(Fmax – Fmin) / 10
Remark:
MAXimum, MINimum are not supported.
SWEep:TIME value
SWEep:TIME:AUTO value
SWEep:TIME:LLIMit value
SWEep:TRIGger:SOURce value

SYSTem Subsystem

Status
Command / Function / Remark / Limitation
SYSTem:ALTernate value
SYSTem:ALTernate:STATe value
SYSTem:COMMunicate:GPIB:ADDRess value
Set / Query the GPIB address of the instrument.
Value:
00 ... 30 | MINimum | MAXimum
Remark:
MINimum, MAXimum are not supported.
SYSTem:DUMP:PRINter?
SYSTem:ERRor?
Query and clear the oldest error number and error description
Remark:
Returns in error in the SM* naming convention
SYSTem:KEY:ASSign value
SYSTem:KEY:CLEar value
SYSTem:KEY:DISable value
SYSTem:KEY:ENABle value
Table 86: Sweep subsystem
94
Rohde & Schwarz

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smb100aSmf100a

Table of Contents