Instrument Initialisation - R&S SMA100A Application Notes

Remote emulation modes with signal generators
Table of Contents

Advertisement

1GP71_0E
AN680XX emulation mode
Status
Command / Function / Remark / Limitation
UNIT:FREQuency value
Select / Query the global default frequency related parameter.
Value:
HZ, KHZ, MHZ, GHZ
Default:
HZ
UNIT:TIME value
Select / Query the global default time related parameter.
Value:
S, MS, US, NS
Default:
S

Instrument initialisation

The (factory) default setting of the AN680XX is applied when
applying a DCL, SDC to the GPIB interface.
o
ransmitting a *RST command
o
transmitting the first command after activating this emulation mode.
o
During initialisation, the AN680XX applies and activates the setting, that was
active just before power down.
The SM* however applies and activates the (factory) default setting of the
AN680XX.
This different behaviour has to the be taken in consideration.
It is strongly recommended to explicitly setup all necessary parameters
by applying the corresponding command.
The following table shows the (factory) default setting, applied by the SM*'s
emulation software.
Subsystem
ABORT
CONTROL
DIAGNOSTIC
DISPLAY
INITIATE
OUTPUT
Table 48: UNIT subsystem
-
-
-
TEXT STATE = ON
-
STATE = ON
54