Increments; Common - R&S SMA100A Application Notes

Remote emulation modes with signal generators
Table of Contents

Advertisement

1GP71_0E
RC3102 emulation mode

Increments

Status
Command / Function / Remark / Limitation
IN0
Enable spin wheel.
IN1
Disable spin wheel.
IN2
Select step size to coarse (fixed).
IN3
Select step size to medium (fixed).
IN4
Select step size to fine (fixed).
IN5
Set step size to the value specified by the FS or AS command.

Common

Status
Command / Function / Remark / Limitation
RS value
Set status byte mask.
GS1
Turn Standby mode on.
Remark:
Standby is simulated, RF output is turned off.
GS0
Turn Standby mode off.
Remark:
RF output is restored to the state before entering standby mode.
SF value
Execute special function.
Value:
80
81
82
83
84
85
Remark:
Only special functions noted above are supported.
DG value
Execute special function.
Remark:
Table 164: Increment commands
transfer cw frequency to sweep start frequency
transfer cw frequency to sweep stop frequency
set sweep dwell time to 2(10) ms
set sweep dwell time to 20 ms
set sweep dwell time to 200 ms
set sweep dwell time to 1 s
170