R&S SMA100A Application Notes page 48

Remote emulation modes with signal generators
Table of Contents

Advertisement

AN680XX emulation mode
Status
Command / Function / Remark / Limitation
[:SOURce:]POWer:ATTenuation:AUTO value
Set / Query the coupling of the attenuation to the amplitude level control circuit.
Value:
ON | OFF | 1 | 0
Default:
ON
[:SOURce:]POWer:ATTenuation:STEP[:INCRement] value
Set / Query the output attenuator step size
Value:
0 ... 110 [DB]
Default:
10.0 dB
[:SOURce:]POWer:CENTer value
Set / Query the RF output center power level
Value:
powermin ... powermax [DBM]
Default:
0.5 * (powermin + powermax)
[:SOURce:]POWer:DISPlay:OFFSet value
Set / Query the offset for the power level display.
Value:
-100 ... +100 [DB]
Default:
0.0 dB
[:SOURce:]POWer:DISPlay:OFFSet:STATe value
[:SOURce:]POWer:MODe value
Set / Query the power level operation mode.
Value:
CW | FIXed | SWEep[1] | SWEep2 | ALSW | LIST[1] | LIST2 | LIST3 | LIST4
Default:
CW
Remark:
SWEep[1], ALSW, LIST<n> are not supported.
SWEep[1] is used for frequency sweep, SWEep2 is used for power sweep.
[:SOURce:]POWer:SLOPe value
[:SOURce:]POWer:SLOPe:PIVot value
[:SOURce:]POWer:SLOPe:STATe value
[:SOURce:]POWer:SLOPe:STEP[:INCRement] value
1GP71_0E
48
Rohde & Schwarz

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smb100aSmf100a

Table of Contents