R&S SMA100A Application Notes page 75

Remote emulation modes with signal generators
Table of Contents

Advertisement

HP8340 emulation mode
R1
Select Active mode.
Remark:
The implementation on SM* is identical to R3.
R2
Set Rf output off.
R3
Set Rf ouput on.
R5
Set Rf output dead.
Remark:
The implementation on SM* is identical to R2.
ST value
Save actual setup to backup register.
Value:
0 ... 9
Remark:
The contents of the registers 0 ... 9 are NOT persisent and are lost when power off
or leaving the emulation mode.
RC value
Recall setup from backup register.
Value:
0 ... 9
Remark:
The contents of the registers 0 ... 9 are NOT persisent and are lost when power off
or leaving the emulation mode.
SV value
Save actual setup to backup register.
Value:
0 ... 99
Remark:
The contents of the registers 0 ... 99 are NOT persisent and are lost when power off
or leaving the emulation mode.
RL value
Recall actual setup from register value.
Value:
0 ... 99
Remark:
The contents of the registers 0 ... 99 are NOT persisent and are lost when power off
or leaving the emulation mode.
SQ
Step forward to next register and recall setup.
GT
Setup a flexible sequence.
1GP71_0E