R&S SMA100A Application Notes page 148

Remote emulation modes with signal generators
Table of Contents

Advertisement

HP8657 emulation mode
R2
Set Rf output off.
R3
Set Rf ouput on.
R5
Set Rf output dead.
Remark:
The implementation on SM* is identical to R2.
ST value
Save actual setup to backup register.
Value:
0 ... 9
Remark:
The contents of the registers 0 ... 9 are NOT persisent and are lost when power off
or leaving the emulation mode.
RC value
Recall setup from backup register.
Value:
0 ... 9
Remark:
The contents of the registers 0 ... 9 are NOT persisent and are lost when power off
or leaving the emulation mode.
SV value
Save actual setup to backup register.
Value:
0 ... 99
Remark:
The contents of the registers 0 ... 99 are NOT persisent and are lost when power off
or leaving the emulation mode.
RL value
Recall actual setup from register value.
Value:
0 ... 99
Remark:
The contents of the registers 0 ... 99 are NOT persisent and are lost when power off
or leaving the emulation mode.
SQ
Step forward to next register and recall setup.
GT
Setup a flexible sequence.
QS
Step backward to previuous register and recall setup.
RP
Reset reverse power protection.
Table 146: Other commands
1GP71_0E
148
Rohde & Schwarz