Lf Control; Lf Generator Frequency - R&S SMA100A Application Notes

Remote emulation modes with signal generators
Table of Contents

Advertisement

1GP71_0E
AF2040 emulation mode
Status
Command / Function / Remark / Limitation
SEQT:PARAM:TDUR value
SEQT:PARAM:TGAP value
SEQT:EDIT:TNUM value
SEQT:EDIT:TFRQ value
SEQT:EDIT:TDUR value
SEQT:EDIT:TGAP value
SEQT:EDIT:LOAD value
SEQT:EDIT:SAVE value
SEQT?
SEQT:MODE?
SEQT:PARAM?
SEQT:EDIT?

LF Control

Status
Command / Function / Remark / Limitation
LF:ON
Turn LF output on.
LF:OFF
Turn LF output off.
LF:GEN
LF:MON value
LF?
Query informations on the LF operation.

LF Generator Frequency

Status
Command / Function / Remark / Limitation
LFGF[:VALUE] value
Set LF generator frequency.
Value:
num [Hz | KHZ | MHZ |GHZ]
Default:
1.0 KHZ
Table 21: Sequential Tones
Table 22: LF Control