HP8340 emulation mode
MD0
Turn off delta marker.
MC
Transfer frequency value of active marker to center frequency.
Remarks:
Affect on center/delta frequency sweep mode depends on firmware release of
HP8340.
SM* always selects center/delta frequency sweep mode.
MP1
Turn on marker 1 ] marker 2 frequency sweep mode.
MP0
Turn off marker 1 ] marker 2 frequency sweep mode.
SHMP
Transfer marker 1 frequency to start frequency and marker 2 frequency to stop
frequency.
Remarks:
Affect on start/stop frequency sweep mode depends on firmware release of
HP8340.
SM* always selects start/stop frequency sweep mode.
AK1
Turn on amplitude frequency markers.
AK0
Turn off amplitude frequency markers.
Table 53: Frequency Marker commands
•
Sweep Mode, Trigger and Time
Status
Command / Function / Remark / Limitation
S1
Select continuous sweep mode.
Remarks:
Default on PON
S2
Select single sweep mode.
SG
Select single sweep mode.
S3
Select stepped (manual sweep) mode.
Remarks:
SM* output signal at connector X AXIS is always in the range from 0V (start
frequency) to 10V (stop frequency).
SM
Select stepped (manual sweep) mode.
Remarks:
SM* output signal at connector X AXIS is always in the range from 0V (start
1GP71_0E