Facility Pre-Installation Summary
12.4.2

Overview of Wire Diameters in Grounding and Functional Cables

Installation Suggestion
This overview provides suggestions on how to fulfill the minimum ExacTrac requirements from the
previous chapters. Local regulatory requirements and your specific needs can lead to a different
solution for implementing wire diameters for supply lines, PE cables and interlock cables. The
LAN cable must be provided for the appropriate Ethernet speed class.
LINAC
110
Bunker Grounding
reference bar.
Grounding reference bar.
Maximum at 3 meters from
Computer Cabinet.
Computer
Cabinet
[1.0]
In-Room Power
Distribution Box
[1.3]
Bunker
Door
Warning
Interlock
Lights
12volts
Off
Customer supplied incl. cabling to
connection point at Computer- and
Generator Cabinet X-ray plus 3 m
Generator
Cabinet
[2.0a]
Control Room
Power Distribution Box
[1.5]
Figure 45
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x
Bunker
Warning
Lights
12volts
On
On
Off
Generator
Cabinet
[2.0b]
Floor Box 2
Floor Box 1
[2.1a]
[2.1b]