Selecting the therapy type and the application mode

Savina 300 offers a selection between the therapy
types ventilation and O
The Tube application mode is used for the
ventilation of intubated patients. The NIV
application mode is used for non-invasive
ventilation.
The therapy type and the application mode can
only be changed in standby mode.
Opening the dialog window
 Press the
Start/Standby key.
Savina 300 opens the Start/Standby (A) page.
A
B
C
B Tube application mode
C NIV application mode
D O
therapy
2
Selecting O
therapy
2
Prerequisite: Savina 300 is in standby mode.
1 Touch the O
therapy button (D).
2
2 Confirm with the rotary knob.
3 Use O
therapy, see page 97.
2
Instructions for use Savina 300 SW 5.n
therapy.
2
D
Selecting the application mode for
ventilation
WARNING
Risk of patient injury
If the alarm limits and ventilation settings are
not adjusted after changing from application
mode NIV to Tube, Savina 300 cannot fully
monitor the ventilation.
Check alarm limits and ventilation settings
and adjust if necessary.
CAUTION
Risk of patient injury
In the NIV application mode, Savina 300 cannot
monitor intubated patients adequately.
For intubated patients, use the application mode
Tube.
CAUTION
Risk of patient injury
If a tube without a cuff is used, or if the cuff is not
inflated, very high leakages may occur.
Activate the leakage compensation, see
"Configuring the ventilation functions"
on page 134.
Select an application mode
Prerequisite: Savina 300 is in standby mode.
1 Touch the button for the corresponding

application mode.

2 Confirm with the rotary knob.
Getting started
79