TotalAlert Infinity™ HTM Medical Gas Notification System
Figure 191: Website Setup Page
Electronic Notification
This setup page is used to configure electronic no-
tification (e-mail, pager, phone text message). The
alarm panel acts as an SMTP client. An SMTP server is
required for electronic notification to function. The To-
talAlert Infinity™ website has the ability to setup email
notification for all messages, all gases, or specifically up
to 3 gases.
By setting up the Electronic Notification, key personnel
can receive notifications of all alarm events.
An SMTP e-mail server must be configured in order for
Electronic Notifications to function. Click "Set Up Email
Server" to configure the server (Figure 4.179).
1.
Click Electronic Notification to access the Electronic
Notification setup page (Figure 4.180).
2.
Enter an e-mail address. Up to twelve e-mail
addresses can be set up.
3.
Choose All Messages, All Gases, or Specific Gases
(up to 3) from pull down menu.
4.
Click Submit.
Figure 192: Electronic Notification Setup Page
4107 9016 60.01
BACnet Setup
The BeaconMedæs TotalAlert Infinity™ HTM control
systems utilize the TCP/ IP family of protocols for
communication and each alarm has a unique IP
address.
Each BACnet enabled TotalAlert Infinity™ alarm
conforms to the BACnet protocol.
•
Protocol Implementation Conformance
Each BACnet enabled TotalAlert Infinity™ alarm has
available a Protocol Implementation Conformance
Statement (PICS). The PICS is a written document
created by BeaconMedæs that identifies the particular
options specified by BACnet that are implemented in
the TotalAlert Infinity™ alarm.
•
BACnet Conformance
BACnet Vendor ID: 543
Alarm Profile: BACnet Smart Sensor (B-SS)
BIBB (BACnet Interoperability Building Blocks)
Supported:
•
Read Property - B (DS-RP-B)
•
Dynamic Device Binding - B (DM-DDB-B)
•
Dynamic Object Binding - B (DM-DOB-B)
Object Type Support: Analog value, Binary value,
Device.
Data Line Layer Options: ISO 8802-3 ("Ethernet") LAN
Character Set Support: ISO 10646 (UTF-8) Encoding.
1.
Click BACnet Setup to access BACnet Setup page
(Figure 4.157).
2.
Enter BACnet Device ID#.
3.
Enter BACnet Device Name.
4.
Choose to Enable or Disable BACnet
Communication.
5.
Click Submit
Figure 193: BACnet Setup Page
4-54