TotalAlert Infinity™ HTM Medical Gas Notification System
1.5.3 Gas Combination Alarms
Figure 1.13: Combination Alarm Front Panel
Gas Combination alarms include a mix of the following
modular components:
•
B30 board
•
B60 board
•
B50 board
Combination alarms can monitor up to 20
BeaconMedaes Line Contact Monitoring signal.
Inputs can be assigned to any one of 8 gas service
indicators.
Combination alarms can also monitor up to 6 digital
gas sensors.
Model Number Scheme:
____ - _______ - ____
Series - Configuration - Retrofit
See Model Number Chart in Section 1.7.
Example:
TotalAlert Infinity™ Combination Alarm
Example Model Number: A3-C10
1.6 Electromagnetic Compatibility
1.6.1 EN 60601-1-2
Medical Electrical Equipment needs special precautions
regarding EMC and needs to be installed and put into
service according to the EMC information provided in
this manual.
Portable and mobile RF communications equipment
can affect Medical Electrical Equipment. The use of
accessories, transducers, and cables other than those
specified by the manufacturer, may result in increased
emissions or decreased immunity of the
TotalAlert Infinity.
The TotalAlert Infinity should not be used adjacent
to, or stacked with, other equipment. If adjacent or
stacked use is necessary, the TotalAlert Infinity should
be observed to verify normal operation in the
configuration in which it will be used.
Guidance and manufacturer's declaration -
electromagnetic emissions
The TotalAlert Infinity is intended for use in the electromag-
netic environment specified below. The customer or the user
of the TotalAlert Infinity should assure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
1-7
Electromagnetic
Compliance
environment - guidance
The TotalAlert Infinity
uses RF energy only for
its internal function.
Therefore, its RF
Group 1
emissions are very low
and are not likely to
cause any interference in
nearby electronic
equipment.
The TotalAlert Infinity
Class A
is suitable for use in all
establishments other
Class A
than domestic and those
directly connected to the
public low-voltage power
supply network that sup-
Complies
plies buildings used for
domestic purposes.
4107 9016 60.01