TotalAlert Infinity™ HTM Medical Gas Notification System
NOTE
:
To learn the IP address, touch the MISCEL-
LANEOUS tab on the configuration screen
then touch the WIRED ETHERNET tab.
Figure 258
Login to Setup Pages
1. Click Login in Setup Section.
2. The web browser will request a user-
name and password (See Figure 259).
The factory defaults are:
Username: TotalAlert
Password: 121212
NOTE
:
If the on-screen password has been changed
from the default password, use the newly
assigned password.
Figure 259
3. The left side bar will now contain the "La-
belSetup" link. (See Figure 260)
4. Click on Label Setup to enter the main
label screen for non-latin language input.
NOTE
:
Logging in applies only to the computer on
which the login occurs. If more than one
PC is being used to access the alarm web
pages, each one must be logged in individu-
ally. The computer stays logged in until the
user closes the browser or until the device
has been rebooted.
The TotalAlert Infinity Central alarm is
designed to allow the setup of 5 Channels
(source badges) with 4 conditions per chan-
nel, for a total of 20 signals.
Before starting, have a complete listing of
Gas-Signal-Location (See Section 9.5).
1. Choose Central Badge Setup button. (See
Figure 261)
9-7
Figure 260
4107 9016 60.01