TotalAlert Infinity™ HTM Medical Gas Notification System
F = Low Alarm Settings Tab
You can remove the low alarm set point or customize
the low alarm set point.
To remove the low alarm set point:
1.
Touch the NONE tab.
2.
Touch the check on the CONFIRM SETTINGS
CHANGE screen.
To customize the low alarm set point:
1.
Touch the ADJUST tab.
2.
Enter the new set point.
3.
Touch the Save button.
Figure 119: Low Alarm Settings
G = High Alarm Settings Tab
You can remove the high alarm set point or customize
the high alarm set point.
To remove the high alarm set point:
1.
Touch the NONE tab.
2.
Touch the check on the CONFIRM SETTINGS
CHANGE screen.
To customize the high alarm set point:
1.
Touch the ADJUST tab.
2.
Enter the new set point.
3.
Touch the Save button.
Active Alarm
To setup active alarm General Message:
1.
Touch the General Message tab to activate general
message.
2.
Touch the gray text box at bottom of screen to type
in general message.
3.
Using the keyboard, enter the message that you
would like to be active when in alarm mode and
save.
To setup active alarm Option Message:
1.
Touch the Optional Message tab to activate
optional message.
2.
Touch the 01 tab.
3.
Using the keyboard, enter the optional message
that you would like to be active when in alarm
mode and save.
Configuring Additional 4-20mA Devices
Configure additional 4-20mA devices by touching the
number on the numeric display that corresponds with
the gas badge and repeating the steps listed above.

4.7.2 Relay Output Board (B4X Board)

To configure a Relay Output, start on the CONFIGURA-
TION screen. Users are always prompted to enter the
password when accessing this screen.
Type the password, and touch the Enter button.
The CONFIGURATION screen displays.
Figure 120: Configuration Screen
4-35