Example: with a filter depth of "5", five measurement cycles must be consistent before an evaluation is
performed. For further information, see also the description of interference suppression (see
chapter 4.12).
For all digital process interfaces, the configuration can also be performed via the respective field-
bus interface (see chapter 9 et seq.) or via the
chapter 12).
When configuring the filter depth, you can enter values of up to 255. Values above 255 (to 299)
are not accepted and must be entered again.
The structure of the configuration in the receiver control panel menu is as follows:
Level 0
Level 1
Main Settings
Commands
Operational setting
 Select Main Settings > Operational Settings > Filter Depth.
8.5.4

Defining the display properties

With these configurations for the display, the brightness and a time unit for darkening the display are
defined.
Visibility:
• Off: no display; the display remains dark until a button is pressed.
• Dark: text is only slightly visible.
• Normal: text is visible with good contrast.
• Bright: text appears very bright.
• Dynamic: The display darkens gradually over the number of seconds configured under Time Unit (s).
During this time span, the display passes through all levels, from bright to off.
After approx. 5 minutes without button actuation, configuration mode is exited and the display
changes to the previous mode.
When configuring the Visibility in the dark, normal and bright modes, the display is completely
inverted after approx. 15 minutes to prevent the LEDs from burning in.
When configuring the Time Unit (s), you can enter values of up to 240 seconds. Values above
240 (to 299) are not accepted and must be entered again.
The structure of these configurations in the receiver control panel menu is as follows:
Level 0
Level 1
Display
Language
Operating
mode
Visibility
Time Unit (s)
Leuze electronic
Level 2
Description
Teach
Filter Depth
(enter value)
min = 1
max = 255
Level 2
Description
English
Process mode
Off
(enter value)
min = 1
max = 240
CML 720i Ex
Starting up the device - Basic configuration
Sensor Studio
configuration software (see
Reset
Factory settings
German
French
Italian
Alignment
Dark
Normal
Bright
Spanish
Dynamic
64