Chapter 10 Troubleshooting; Overview Of Troubleshooting; Read And Troubleshoot Fault And Warning Messages - Emerson Rosemount FCL-01 Reference Manual

Free chlorine system with rosemount 56 transmitter
Hide thumbs Also See for Rosemount FCL-01:
Table of Contents

Advertisement

Reference Manual
00809-0200-3412
10 
Troubleshooting
10.1 

Overview of troubleshooting

When the transmitter identifies a problem the word warning or fault appears
intermittently at the bottom of the display. To read a fault or warning message and
troubleshooting information, press INFO.
See
Read and troubleshoot Fault and Warning
Warning
Fault
10.2 
Read and troubleshoot Fault and Warning
messages
Complete the following steps to access Fault and Warning messages and determine what
they mean.
Procedure
1. From the main display, press INFO.
Rosemount FCL 56
The instrument or sensor is usable, but you should take steps as soon as
possible to correct the condition causing the warning. Warning messages can
be turned off. To turn off warning messages, go to the main menu and choose
Display setup. Scroll to the Warning tab and turn off warning messages.
The measurement is seriously in error and is not to be trusted. A fault condition
might also mean that the transmitter has failed. Correct fault conditions
immediately. When a fault occurs, the analog output goes to 22.00 mA or the
value programmed in Outputs. Fault messages cannot be turned off.
The following screen appears.
messages.
Troubleshooting
October 2023
51