Measurement support
Figure 9-8:
D3 or D3 and D4 density calibration using the Field Communicator
Close shutoff valve
downstream from sensor
9.9
Perform temperature calibration
Temperature calibration establishes the relationship between the temperature of the
calibration fluids and the signal produced by the sensor.
142
D3 Calibration
Fill sensor with D3 fluid
On-Line Menu >
Service Tools >
Maintenance >
Density Calibration
Dens Pt 3 T-Series
Calibration Method
executes
Enter density of D3 fluid
OK
Calibration in Progress
message
Density Calibration Complete
message
OK
Home
Micro Motion Model 1700 Transmitters with Analog Outputs
D4 Calibration
Fill sensor with D4 fluid
Service Tools >
Maintenance >
Density Calibration
Dens Pt 4 T-Series
Calibration Method
executes
Enter density of D4 fluid
OK
Calibration in Progress
message
Density Calibration Complete
message
OK
Home
Done