Configure process measurement

Topics covered in this chapter:
4.1

Configure mass flow measurement

The mass flow measurement parameters control how mass flow is measured and reported.
4.1.1

Configure Mass Flow Measurement Unit

Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Unit
Overview
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass
flow rate. The unit used for mass total and mass inventory is derived from this unit.
Any selected measurement unit, (mass, volume or gas standard volume), is automatically
applied to both the mA and Frequency Outputs.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Configuration and Use Manual
Configure mass flow measurement
Configure volume flow measurement for liquid applications
Configure GSV flow measurement
Configure Flow Direction
Configure density measurement
Configure temperature measurement
Configure pressure compensation
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > MASS
Device Tools > Configuration > Process Measurement > Flow
Configure process measurement
17