Quick start
2.2

Check meter status

Check the meter for any error conditions that require user action or that affect
measurement accuracy.
1.
2.
Related information
2.2.1

Transmitter status reported by LED

Table 2-1:
Transmitter status reported by status LED
LED state
Description
Solid green
No alerts are active.
Flashing green (if ena-
Unacknowledged corrected condition (no
bled)
alert)
Solid yellow
One or more low-severity alerts are active.
Flashing yellow (if ena-
Calibration in progress, or Known Density Ver-
bled)
ification in progress.
One or more low-severity alerts are active and
have not been acknowledged.
Solid red
One or more high-severity alerts are active.
Flashing red (if ena-
One or more high-severity alerts are active
bled)
and have not been acknowledged.
If Status LED Blinking is disabled, all LEDs will show a solid color rather than flashing.
6
Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alert A009 is active.
This alert should clear automatically when the power-up sequence is complete.
Check the status LED on the transmitter.
View and acknowledge status alerts
Recommendation
Continue with configuration or process meas-
urement.
Continue with configuration or process meas-
urement. Acknowledge the alert if you choose.
A low-severity alert condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement, but Micro Motion still rec-
ommends identifying and resolving the alert
condition.
A low-severity alert condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement, but Micro Motion still rec-
ommends identifying and resolving the alert
condition.
A high-severity alert condition affects meas-
urement accuracy and output behavior. Re-
solve the alert condition before continuing.
A high-severity alert condition affects meas-
urement accuracy and output behavior. Re-
solve the alert condition before continuing.
Acknowledge the alert if you choose.