Modification type
Adjustment
Configuration and Use Manual
Change
Sensors that are not straight tube sensors are now correctly
identified
The mA Output fixed alert is now set
The Factory Configuration Invalid status bit is now set cor-
rectly when connected to a 700 core processor — as the 700
core processor does not support saving and restoring the fac-
tory configuration
Enabling a new feature on a flow-only device does not reset
the output configuration to factory defaults
Polled variables are now converted to the transmitter units
before being used
HART Primary, Secondary, and Tertiary Variables can be
changed using HART when the transmitter is configured for
gas standard volume
Petroleum measurement variables are removed from the dis-
play when petroleum measurement is disabled
The mA Output can be fixed when assigned to a concentra-
tion measurement, petroleum measurement, or GSV variable
when the transmitter is not connected to a core processor
The units for concentration curve data are correctly conver-
ted when the data is entered using HART
Meter verification failed and aborted alerts are now reflected
in the HART More Status flag
The HART Squawk command is functional when the HART
write protect is enabled
Locking the transmitter, unlocking the transmitter, or chang-
ing write protection does not set the HART Configuration
Changed flag
If connected to a 700 core processor with software version
3.1 or earlier, there is no longer a core processor write failed
alert when setting volume flow units to beer barrels per time
unit
The factory zero value is immediately updated from the core
processor after running a zero calibration and saving the fac-
tory configuration
A firmware upgrade done without clearing non-volatile mem-
ory no longer causes an EEPROM Failure alert
The status bit for the undefined A141 alert (Device Data Cap-
ture Triggered) no longer appears in the Modbus status regis-
ter
When using AMS to configure the transmitter:
-
Changing the output configuration for mA 1, mA 2, or the
Frequency Output, does not change the configuration of
all three outputs
-
mA configuration methods can be cancelled
-
Density and temperature units display on the concentra-
tion measurement configuration page
NE 53 history