Table 5-2:
Status alerts and Status Alert Severity
Alert code
Status message
A014
Transmitter Failure
A016
Sensor RTD Failure
A017
T-Series RTD Failure
A018
EEPROM Error (Transmit-
ter)
A019
RAM Error (Transmitter)
A020
No Flow Cal Value
A021
Incorrect Sensor Type (K1)
A022
Configuration Database
Corrupt (Core Processor)
A023
Internal Totals Corrupt
(Core Processor)
A024
Program Corrupt (Core
Processor)
A025
Boot Sector Fault (Core
Processor)
A026
Sensor/Transmitter Com-
munications Failure
A028
Core Processor Write Fail-
ure
A031
Low Power
A032
Meter Verification in Pro-
gress: Outputs to Fault
A033
Insufficient Right/Left Pick-
off Signal
A034
Meter Verification Failed
Meter Verification Aborted Fault
A035
A100
mA Output 1 Saturated
A101
mA Output 1 Fixed
A102
Drive Overrange
Configuration and Use Manual
(continued)
Default severity Notes
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Applies only to flowmeters with the
standard core processor.
Fault
Applies only to flowmeters with the
standard core processor.
Fault
Applies only to flowmeters with the
standard core processor.
Fault
Applies only to flowmeters with the
standard core processor.
Fault
Fault
Fault
Applies only to flowmeters with the
enhanced core processor.
Varies
Applies only to transmitters with
Smart Meter Verification.
If outputs are set to Last Measured
Value, severity is Info. If outputs are
set to Fault, severity is Fault.
Fault
Applies only to flowmeters with the
enhanced core processor.
Fault
Applies only to transmitters with
Smart Meter Verification.
Applies only to transmitters with
Smart Meter Verification.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Configure device options and preferences
Configurable?
No
Yes
Yes
No
No
Yes
No
No
No
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
65