Emerson Micro Motion 1700 Configuration And Use Manual page 72

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 1700:
Table of Contents

Advertisement

Configure device options and preferences
Status alerts and options for Status Alert Severity
Table 5-2:
Status alerts and Status Alert Severity
Alert code
Status message
A001
EEPROM Error (Core Pro-
cessor)
A002
RAM Error (Core Processor) Fault
A003
No Sensor Response
A004
Temperature Overrange
A005
Mass Flow Rate Overrange
A006
Characterization Required
A008
Density Overrange
A009
Transmitter Initializing/
Warming Up
A010
Calibration Failure
A011
Zero Calibration Failed:
Low
A012
Zero Calibration Failed:
High
A013
Zero Calibration Failed:
Unstable
64
Option
Description
Fault
Actions when fault is detected:
The alert is posted to the Alert List.
Outputs go to the configured fault action (after Fault Timeout has expired, if
applicable).
Digital communications go to the configured fault action (after Fault Time-
out has expired, if applicable).
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
Outputs return to normal behavior.
Digital communications return to normal behavior.
The status LED (if available) returns to green and may or may not flash.
Informa-
Actions when fault is detected:
tional
The alert is posted to the Alert List.
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
The status LED (if available) returns to green and may or may not flash.
Ignore
No action
Default severity Notes
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Micro Motion Model 1700 Transmitters with Analog Outputs
Configurable?
No
No
Yes
No
Yes
Yes
Yes
Yes
No
Yes
Yes
Yes

Advertisement

Table of Contents
loading

Table of Contents