Troubleshooting
Alert num-
ber
Alert title
A114
mA Output 2 Fixed
A115
No External Input or
Polled Data
A118
Discrete Output 1
Fixed
A131
Meter Verification in
Progress: Outputs to
Last Measured Value
A132
Sensor Simulation
Active
N/A
Density FD Calibra-
tion in Progress
N/A
Density D1 Calibra-
tion in Progress
N/A
Density D2 Calibra-
tion in Progress
N/A
Density D3 Calibra-
tion in Progress
N/A
Density D4 Calibra-
tion in Progress
N/A
Zero Calibration in
Progress
N/A
Reverse Flow
158
Possible cause
The mA Output is configured to
send a constant value.
The connection to an external
measurement device has failed. No
external data is available.
The Discrete Output has been con-
figured to send a constant value.
A meter verification test is in pro-
gress, with outputs set to Last Meas-
ured Value.
Sensor simulation is enabled.
A flowing density calibration is in
progress.
A D1 density calibration is in pro-
gress.
A D2 density calibration is in pro-
gress.
A D3 density calibration is in pro-
gress.
A D4 density calibration is in pro-
gress.
A zero calibration is in progress.
Flow through the device is in the re-
verse direction (against the flow ar-
row).
Micro Motion Model 1700 Transmitters with Analog Outputs
Recommended actions
•
Check whether the output is in loop test
mode. If it is, unfix the output.
•
Exit mA Output trim, if applicable.
•
Check whether the output has been set
to a constant value via digital communi-
cation.
•
Verify that the external device is operat-
ing correctly.
•
Verify the wiring between the transmit-
ter and the external device.
•
Verify the HART polling configuration.
•
Check whether the output is in loop test
mode. If it is, unfix the output.
•
Allow the procedure to complete.
•
Disable sensor simulation.
•
No action required.
•
No action required.
•
No action required.
•
No action required.
•
No action required.
•
No action required.
•
No action required.