Microalbumin/Creatinine Test; Dca Vantage Operator's Guide; Preparing Patient Samples And Controls For The Microalbumin/Creatinine Test; Performing A Dca Microalbumin/Creatinine Test - Siemens DCA Vantage Operator's Manual

Table of Contents

Advertisement

5. Pull the reagent cartridge out of the compartment.
CAUTION
When handling the reagent cartridge, do not force the removal of the
cartridge. This can cause physical injury to you and damage to the
system.
6. Close the system door.
7. Discard the cartridge in a proper container, according to your standard
laboratory procedures.

Microalbumin/Creatinine Test

The following sections explain the test sequence for a Microalbumin/
Creatinine test.
Preparing Patient Samples and Controls for the Microalbumin/Creatinine
Test
Materials Required:
DCA Systems Microalbumin/Creatinine Reagent Kit (For use with the
DCA Vantage system)
patient sample
Creatinine Control Kit or other control
lint-free tissue
For information on preparing controls, refer to Quality Control‚ page 87.
For information on calibrating, refer to Calibrating for Microalbumin/Creatinine‚
page 85.

Performing a DCA Microalbumin/Creatinine Test

BIOHAZARD
Wear personal protective equipment, including safety glasses and
gloves. Use universal precautions. Refer to Safety Information‚
page 167 for recommended precautions when working with
biohazardous materials.
You can save unused capillary holders and use them with any lot of DCA
Systems Microalbumin/Creatinine reagent cartridges.
1. Open and inspect the contents of the Reagent Cartridge foil package.
Refer to Opening and Inspecting the Reagent Foil Package‚ page 45.

DCA Vantage Operator's Guide

63

Advertisement

Table of Contents
loading

Table of Contents