Error Code
Description
E109
The absorbance of one
of the agglutinations is
HbA
1c
too low.
Cartridge error
Possible cause:
- reading for
glycated
hemoglobin is
too low
E110
The absorbance of one
of the agglutinations is
HbA
1c
too high.
Cartridge Error
Possible cause:
- reading for
glycated
hemoglobin is
too high
E111
The delta in absorbance
between the first and the
HbA
1c
last agglutination
Cartridge Error
readings is too low.
- low glycated
Possible cause:
hemoglobin
response
E112
The delta in absorbance
between the first and
HbA
1c
last agglutination
Cartridge Error
reading is too high.
- very large
Possible cause:
glycated
hemoglobin
response
DCA Vantage Operator's Guide
cartridge exposed to
excessive humidity
or temperature
cartridge exposed to
excessive humidity
or temperature
cartridge exposed to
excessive humidity
or temperature
cartridge exposed to
excessive humidity
or temperature
Corrective Action
1. Discard the sample.
2. Acknowledge the error.
The system is ready.
3. Repeat the test with a new
sample.
Verify that the reagent kit
NOTE:
was stored properly.
4. If the error still occurs, contact
your local technical support.
1. Discard the sample.
2. Acknowledge the error.
The system is ready.
3. Repeat the test with a new
sample.
Verify that the reagent kit
NOTE:
was stored properly.
4. If the error still occurs, contact
your local technical support.
1. Discard the sample.
2. Acknowledge the error.
The system is ready.
3. Repeat the test with a new
sample.
Verify that the reagent kit
NOTE:
was stored properly.
4. If the error still occurs, contact
your local technical support.
1. Discard the sample.
2. Acknowledge the error.
The system is ready.
3. Repeat the test with a new
sample.
NOTE:
Verify that the reagent kit
was stored properly.
4. If the error still occurs, contact
your local technical support.