Setting User Corrections

You can adjust the slope and the offset for HbA
Creatinine values.
Corrected and uncorrected results outside the operating range are reported
the same way.
If results are within the operating range and a user correction is applied, the
corrected results are reported with an asterisk (*).
When recalling saved data, the saved data displays as uncorrected if a
correction is not set in the system. If a correction is set when the data is
recalled, that correction is applied to the recalled data.
Setting the HbA
User Correction
1c
Use the HbA
User Correction screen to adjust the slope and the offset for
1c
HbA
values.
1c
1. At the Home screen, select
2. Select
System Settings
3. At the System Settings menu, select
4. Select
Patient Tests
5. Select
HbA1c User Correction
6. Change the slope, if needed:
a. Select
.
Edit
The HbA
Slope screen displays.
1c
b. Enter the slope.
c. Select
.
Enter
The HbA
User Correction screen displays.
1c
7. Change the offset, if needed:
a. Select
.
Edit
The HbA
Offset screen displays.
1c
b. Enter the offset.
c. Select
.
Enter
The HbA
User Correction Screen displays.
1c
To reset the settings to the default values:
a. Select
Reset Default Values
This resets the slope to 1.000 and the offset to 0.0.
8. After completing the changes, select
The Patient Tests screen displays.
DCA Vantage Operator's Guide
Menu
.
from the Menu screen.
Additional Settings
from the Additional Settings menu.
.
.
Back
values, Albumin values, and
1c
.
.
155