Siemens DCA Vantage Operator's Manual page 122

Table of Contents

Advertisement

System Message
Change the Air Filter.
Clean the Barcode window.
Clean the Cartridge Compartment.
Compressing the database, please
wait.
Data storage is at > 90% of the
storage capacity. Change Storage
setting to AUTO or delete some old
tests.
Enabled ASTM ports are not ready to
transmit data, please check
connections.
HbA1c Abnormal control test failed.
You cannot perform HbA1c patient
tests.
HbA1c Abnormal control test
requested by remote computer.
HbA1c Abnormal control test
requested by remote computer.
Patient test locked until the control
test is run.
HbA1c Normal control test failed. You
cannot perform HbA1c patient tests.
HbA1c Normal control test requested
by remote computer.
HbA1c Normal control test requested
by remote computer. Patient test
locked until the control test is run.
Instrument is Initializing, please wait.
Invalid user action.
122
Comment
A reminder message that it is time to
replace the air filter. The frequency of
the reminder is determined by your
location.
A reminder message that it is time to
clean the onboard barcode reader
window. The frequency of the
reminder is determined by your
location.
A reminder message that it is time to
clean the cartridge area. The
frequency of the reminder is
determined by your location.
Warning message to clear old tests.
Check the Ethernet and serial
connections to ensure cables are
secure.
The control test failed. Rerun the
control test with a new control
cartridge.
DCA Vantage Operator's Guide