Error Code
Description
E309
The %CV of the
Microalbumin
Microalbumin/
transmittance reading is
Creatinine
too high. (Particulate
Cartridge Error
contamination.)
- high variation
in Microalbumin
readings
E310
One of the creatinine
absorbance readings is
Microalbumin/
too low.
Creatinine
Cartridge Error
Possible cause:
- Creatinine
absorbance is
too low
E311
One of the 26 creatinine
absorbance readings is
Microalbumin/
too high.
Creatinine
Cartridge Error
Possible cause:
- Creatinine
absorbance is
too high
E313
The Mean Squared Error
calculated after the third
Microalbumin/
order polynomial fit of
Creatinine
the creatinine readings
Cartridge Error
is too large.
- high variability
in absorbance
Possible cause:
readings
DCA Vantage Operator's Guide
cartridge exposed to
excessive
temperature or
humidity
cartridge exposed to
excessive
temperature or
humidity
cartridge exposed to
excessive
temperature or
humidity
Corrective Action
1. Discard the sample.
2. Acknowledge the error.
The system is ready.
3. Repeat the test.
4. If the error still occurs, contact
your local technical support.
1. Discard the sample.
2. Acknowledge the error.
The system is ready.
3. Repeat the test.
Verify that the reagent
NOTE:
was stored properly, and that the
foil package was opened just prior
to use.
4. If the error still occurs, contact
your local technical support.
1. Discard the sample.
2. Acknowledge the error.
The system is ready.
3. Repeat the test.
Verify that the reagent
NOTE:
was stored properly, and that the
foil package was opened just prior
to use.
4. If the error still occurs, contact
your local technical support.
1. Discard the sample.
2. Acknowledge the error.
3. Repeat the test.
Verify that the reagent was
NOTE:
stored properly, and that the foil
package was opened just prior to
use.
4. If the error still occurs, contact
your local technical support.
145