6. Change the slope, if needed:
a. Select
.
Edit
The Creatinine Slope screen displays.
b. Enter the slope.
c. Select
.
Enter
The Creatinine User Correction screen displays.
7. Change the offset, if needed:
a. Select
.
Edit
The Creatinine Offset screen displays.
b. Enter the offset.
c. Select
.
Enter
The Creatinine User Correction Screen displays.
8. To reset the settings to the default values, select
This resets the slope to 1.000 and the offset to 0.0.
9. Select
.
Back
The Patient Tests screen displays.

Setting the GFR Calculation

Use the GFR Calculation screen to turn the GFR Calculation on or off, and to
select the unit of measurement for plasma creatinine.
1. At the Home screen, select
2. Select
System Settings
3. At the System Settings menu, select
4. At the Additional Settings menu, select
5. Select
GFR Calculation
6. Select one of the following options under GFR Calculation:
Enabled
Disabled
7. Select one of the following units under Creatinine Units:
mg/dL
mmol/L
8. To save the GFR Calculation settings, select

Setting the Patient Recall List Options

Use the Patient Recall List Options screen to set the search criteria for patient
results, and how the results are displayed. You can search by either patient ID
or last name.
DCA Vantage Operator's Guide
.
Menu
from the Menu screen.
Additional Settings
Patient Tests
.
Reset Default Values
.
.
.
Save
.