Setting The Albumin User Correction; Setting The Creatinine User Correction - Siemens DCA Vantage Operator's Manual

Table of Contents

Advertisement

Setting the Albumin User Correction

Use the Albumin User Correction screen to adjust the slope and the offset for
Albumin values.
Siemens recommends that you do not use a user defined correlation
NOTE:
when you perform external proficiency testing. This can impact the results.
1. At the Home screen, select
2. Select
System Settings
3. At the System Settings menu, select
4. At the Additional Settings menu, select
5. Select
Albumin User Correction
6. Change the slope, if needed:
a. Select
.
Edit
The Albumin Slope screen displays.
b. Enter the slope.
c. Select
.
Enter
The Albumin User Correction screen displays.
7. Change the offset, if needed:
a. Select
.
Edit
The Albumin Offset screen displays.
b. Enter the offset.
c. Select
.
Enter
The Albumin User Correction Screen displays.
To reset the settings to the default values:
a. Select
Reset Default Values
This resets the slope to 1.000 and the offset to 0.0.
8. After completing the changes, select
The Patient Tests screen displays.

Setting the Creatinine User Correction

Use the Creatinine User Correction screen to adjust the slope and the offset
for creatinine values.
1. At the Home screen, select
2. Select
System Settings
3. At the System Settings menu, select
4. At the Additional Settings menu, select
5. Select
Creatinine User Correction
156
.
Menu
from the Menu screen.
Additional Settings
Patient Tests
.
.
.
Back
.
Menu
, from the Menu screen.
Additional Settings
Patient Tests
.
.
.
.
.
DCA Vantage Operator's Guide