Optical Tests
If the system is set to restricted, to access System Tests you must
NOTE:
enter an authorized Operator ID and access code, refer to Operator IDs and
Access Codes‚ page 42.

Performing User System Tests

The User System Tests screen enables you to test the DCA Vantage system.
1. At the System Menu screen, select
2. From the System Test menu, select
3. Select the check box of the test you want to perform.
4. To initiate the test, select
The tests that require no additional action automatically run.
Some options may not be available because they were not configured
NOTE:
in your system.
The following table explains the different tests you can perform, if enabled,
and if they require additional action:
If you select . . .
Touchscreen
Sound
Door
Onboard Barcode
Reader
External Barcode
Reader
Onboard Printer
External Printer
Serial Port
DCA Vantage Operator's Guide
System Tests
User System Tests
.
Start
then. . .
Select each numbered key. The key changes from a
number to a check mark. If a section fails, the number
remains.
The test times out if all the buttons are not
NOTE:
selected within 60 seconds.
1. Listen for an alert sound.
2. If you hear an alert sound, select
3. If you do not hear an alert sound, select
Open and close the door to verify that the screen
reflects the status of either the open or closed state.
Scan a DCA barcode with the onboard barcode
reader to verify that the onboard barcode reader
works properly.
Scan a DCA barcode with the external barcode reader
to verify that the external barcode reader works
properly.
Verify that the onboard printer prints a report.
Verify that the external printer prints a report.
After connecting a loopback connector, the system
automatically runs the test.
.
.
.
Yes
.
No
109