Error Code
Description
E315
The Microalbumin
Normalization was not
Microalbumin
performed or the
normalization
Microalbumin
required
Normalization Factor
(MNF) was corrupted.
Possible cause:
E316
The Microalbumin
Normalization Factor
Microalbumin
entered is outside of
normalization
expected range.
out of range
Possible causes:
146
system not adjusted
to run Microalbumin/
Creatinine test
normalization
cartridge is not in
proper position
normalization
cartridge is dirty or
defective
Corrective Action
1. Acknowledge the error.
2. If the error still occurs, contact
your local technical support.
1. Acknowledge the error.
2. Verify that the cartridge was
inserted completely into the
system.
3. Repeat the test.
4. If the error still occurs, contact
your local technical support.
DCA Vantage Operator's Guide