Setting Up Auto Send - Siemens DCA Vantage Operator's Manual

Table of Contents

Advertisement

2. Use the numeric keypad to enter the Port Number.
3. To save the Port Number, select

Setting up Auto Send

When you have enabled communications with an LIS, you can configure the
system to automatically send all patient and control test results to the LIS as
soon as the test result has completed, or as soon as you have finished
entering or modifying patient demographic data. Automatic sending of results
(Auto Send) to an LIS is disabled by default. For more information about Auto
Send, see Automatically Sending Data to an LIS‚ page 202.
To enable Auto Send, perform the following procedure:
1. From the Home screen, select
Settings > Connectivity
2. Select the type of connection,
or
POCT1A
.
3. To enable Auto Send, select the
This selection is disabled by default.
To enable manually sending selected results, clear this check box.
4. Select
Configure
At this stage, the system does not verify whether a connection
NOTE:
exists.
5. Continue configuring the Ethernet or Serial connection parameters.
6. When finished, select
The Auto Send setting is included in the backup and restore data so an
administrator can copy the settings from one DCA system to another.
DCA Vantage Operator's Guide
Save
Menu > System Settings > Additional
.
Ethernet
Automatically Send
to set up the Ethernet or Serial port connection.
.
Save
.
or
, and the protocol,
Serial
check box.
ASTM
37