Quality Control Errors
If the control results fall outside the values stated in the package insert, the
following sources of error may have occurred:
Possible Cause
Deterioration of the
reagent cartridge
test areas due to
exposure to light,
ambient moisture,
or heat.
Deterioration of the
control solution.
DCA Vantage Operator's Guide
Corrective Action
Use a new reagent cartridge to repeat the quality control
procedure.
Use a fresh box of reagent cartridges, or a new lot.
If the new reagent cartridge fails to give results within the
expected values, proceed to the next possible cause.
Use a fresh control solution to repeat the quality control
procedure.