Siemens DCA Vantage Operator's Manual page 94

Table of Contents

Advertisement

6. At the Control Test Reminder screen, select 1 of the following:
HbA
Normal
1c
HbA
Abnormal
1c
A/C Low
A/C High
Once selected, the Schedule button to the right of the control test selected
becomes active.
7. Select
Schedule
8. Select
.
Change
9. Select
Optional
or
10. Select
Custom
.
A screen displays that enables you to select a number between 7 and 60
by pressing up or down arrow keys. The selected number displays above
the arrow keys. The default value is 14.
Both up and down arrow keys cycle continuously through the 7 to 60
number range. For example, if 7 displays, pressing the down arrow key
displays 60. Press the down arrow again to display 59, then 58, and so on.
11. Press an arrow key until the number you want displays.
12. Select
.
Next
13. Select the time in hours and minutes, using the arrow keys.
14. Select
or
AM
PM
15. Select
.
Next
The Settings screen displays. Verify the settings you entered are correct. If
a setting is incorrect, select
16. Select
.
Save
The Control Test Reminders screen displays.
17. Repeat steps 6 to 16 to set additional QC reminders, as needed.