Troubleshooting Reagent Messages; Errors Codes And Corrective Actions - Siemens DCA Vantage Operator's Manual

Table of Contents

Advertisement

Symptom
Touchscreen not
responding or not
responding
correctly.

Troubleshooting Reagent Messages

These message display when an error occurs with the reagent cartridge.
Symptom
Reagent not
absorbed by
cartridge filter at
the end of the
test.
Thermal Dot has
reacted (HbA
1c
reagents only)

Errors Codes and Corrective Actions

This section of the guide lists error codes and messages, along with a
description and corrective actions.
DCA Vantage Operator's Guide
Possible Cause
Action
Touchscreen
Run the touchscreen calibration.
needs to be
Refer to Calibrating the
calibrated.
Touchscreen‚ page 110.
Possible Cause
Action
The
Verify that the system is properly
DCA Vantage
installed on a level surface.
system is not on
a level work
surface.
The filter in the
Contact your local technical support
capillary holder is
provider.
not absorbent.
Product was not
There are 2 indicators: outside and
shipped or stored
inside. The outer shipper has an
properly.
over temperature indicator. If this
indicator is red or black, the product
was not stored properly. Check the
inside indicator to confirm. If the
inside indicator is red, the
cartridges should be discarded. The
outside indicator can be red or
black, while the inside indicator is
normal. In this case, the cartridges
can be used.
129