Chapter 4 - Troubleshooting and Maintenance

Analyzer messages and troubleshooting

The Atellica VTLi Immunoassay Analyzer displays messages to let you know of issues or
errors. Messages typically explain the event, provide possible causes, and offer possible
solutions.
!
Details
OK
An error code also displays on the screen.
For a complete listing of errors, see the Atellica VTLi Advanced User Guide.
Icon
Description
Indicates a problem is detected but you can continue using the analyzer.
Indicates a problem is detected that needs correction before continuing.
Some screens have a Details button you tap to get additional information on the issue.

General Troubleshooting

The following table describes some possible situations you might face and gives actions you
can take to solve the problem.
Issue
Analyzer doesn't switch on.
Analyzer does not respond.
You can...
Dock the analyzer in the docking station. Make sure the
docking station is connected to a power supply. If the
analyzer still won't switch on after docking, press the
Reset button.
• Press the Reset button.
• Check the analyzer battery for proper installation and
connection. If necessary, replace the battery.
• If the analyzer doesn't respond after charging, call
your local administrator for assistance.