Multifors 2 - Parallel Bench-top Bioreactor
Preparations for Use
7.4.7 Polarising a (polarographic) METTLER pO
7.4.8 Calibrating the pO
Page 156 of 211
Polarographic pO
be polarised at initial operation or after disconnection from the volt-
age source. Correct calibration is not possible otherwise. This
means that the pO
Sensor
2
The bioreactor system is preconfigured either for using polaro-
graphic METTLER pO
type Visiferm DO ARC.
Generally it applies that: calibrating the pO
formed after sterilisation because sterilisation can change the gra-
dient of the pO
A 1-point calibration to 100 % is generally sufficient for an exact
measurement and should be carried out again before each fermen-
tation.
The zero point (0 % calibration) should be checked at regular inter-
vals of about 6 months.
A detailed description how to calibrate the pO2 sensor can be
found in the separate touch screen software manual from the
equipment manufacturer.
METTLER pO
polarised before calibration.
Sensor
2
sensors from the manufacturer METTLER must
2
sensor must be polarised before calibration.
2
NOTICE!
Details about polarisation, calibration, use mainte-
nance and servicing can be found in the separate
sensor documentation of the manufacturer.
sensors or HAMILTON pO
2
sensor.
2
sensors are polarographic and therefore need to be
2
NOTICE!
Detailed information on the calibration, use,
maintenance and servicing can be found in the
separate sensor documentation of the manufac-
turer.
sensors of the
2
sensor should be per-
2
28 March 2018