Connecting The Sparger To The Gas Supply - Infors HT Multifors 2 Operating Manual

Parallel bench-top bioreactor
Table of Contents

Advertisement

Multifors 2 - Parallel Bench-top Bioreactor
Preparations for Use

7.4.3 Connecting the Sparger to the Gas Supply

Procedure
Page 150 of 211
The sparger must be connected to the gas supply on the basic unit
after sterilisation. Proceed as follows:
1.
Cut off piece of thick-walled silicone hose.
The hose must be long enough to reach from the sparger to
the gas supply on the basic unit without tension or kinks.
2.
Fit the silicone hose to the hose nozzle below the rotameter
and secure it with a cable tie.
3.
Remove the aluminium foil from the inlet air filter.
4.
Plug the end of the hose onto the inlet air filter.
5.
Open the clamp on the sparger hose line.