28 March 2018
11. Fit the silicone hoses onto the addition port adapters on the
vessel top plate.
12. Fit the silicone hoses of the antifoam pumps onto the mounted
antifoam sensors in the vessels.
13. Ensure a correct connection of the reagent bottles to the
pumps (e.g. acid with acid pump etc.).
14. Hang the reagent bottles onto the vessel holder, using the
slots provided.
The picture on the left shows an empty reagent bottle and
serves for general understanding.
15. Close the clamps on the silicone hoses of the reagent bottles.
The hoses should be clamped off as close to the reagent bot-
tle as possible.
Ensure no corrective reagent can get into the vessel.
16. Lightly cap filters on the reagent bottles with a little aluminium
foil.
Multifors 2 - Parallel Bench-top Bioreactor
Preparations for Use
Page 131 of 211
Need help?
Do you have a question about the Multifors 2 and is the answer not in the manual?
Questions and answers