Multifors 2 - Parallel Bench-top Bioreactor
Preparations for Use
Page 114 of 211
3.
Unhook the threaded bolt.
CAUTION!
Risk of contamination and danger of loss of
property!
If the preset length of 21 mm of the spring on the
threaded bolt of the closed clamping ring is
changed using the screw at the end of the bolt, the
sealing of the vessel top plate cannot be guaran-
teed anymore. This carries a risk of contamination
and loss of property.
Therefore:
– Do not alter the preset length of 21 mm (closed
clamping ring) of the spring on the threaded
bolt.
– Preferably, always use the same clamping ring
with the same vessel.
– Check length of spring, if in doubt.
4.
Take off the clamping ring from the vessel and put aside.
28 March 2018