Multifors 2 - Parallel Bench-top Bioreactor
Technical Data
3.5.11 Pumps
Page 40 of 211
NOTICE!
Details about technical data, use and maintenance
of the sensors are in the separate documentation
from the sensor manufacturers.
Description
Value
Type
Peristaltic
Standard
Digital
(3 pieces)
Analogue
(1 piece)
Option
1 or several free-standing pumps, analogue
Rotation speed
Digital
Analogue
Accuracy
± 1 % FS
Pump hoses & flow rates
Description
Value
Material
PharMed BPT
Standard
iD: 1.0 mm
Wall thickness: 1.1 mm
Flow rate: 3.5 ml min
Option 1
iD: 0.5 mm
Wall thickness: 1.15 mm
Flow rate: 1.2 ml min
Option 2
iD: 2.5 mm
Wall thickness: 1,0 mm
Flow rate: 17.2 ml min
Flow rates
All flow rates at 74 min
Acid
Base
AF = Antifoam
Feed
74 min
/ fixed rotation speed
-1
74 min
/ max. rotation speed,
-1
adjustable within range of 0 % up
to 100 %
-1
-1
-1
-1
(100 % rotation speed)
28 March 2018
Need help?
Do you have a question about the Multifors 2 and is the answer not in the manual?