Care
Reassembling system
Connecting anaesthetic gas scavenging system (AGS)
Connect sample line
Connecting anaesthetic gas scavenging
system (AGS)
1 Insert the transfer hose, complete with scavenging
adapter, into the breathing system from underneath
until it engages, or connect the hose to the
scavenging adapter of the semi open system.
2 Route the extractor hose round the equipment to the
rear and
3 connect it to the connector on the collecting system.
Refer to the separate Instructions for Use of the AGS
anaesthetic gas scavenging system.
Connect sample line
For sampling CO
, anaesthetic agent and O
2
4 Screw filter into T-piece.
5 Insert T-piece into the patient connection of the
Y-piece – filter facing upwards to prevent congestion
by droplets of liquid.
6 Securely screw sample line onto filter and water trap.
Only use original sample lines, others may lead to
changes in the machine's technical specifications!
7 The Y-piece with direct connection for the sample line
can also be used instead of the T-piece and filter.
Alcohol or agents containing alcohol must not be
allowed to enter the sample line!
Alcohol will result in incorrect measurement of the
concentration.