Tabular summary

Machine components
Stripped down basic machine
Parameter box (optional)
Rubber parts:
Tube / mask
Y-piece
Breathing hoses
Breathing bag
Breathing bag hose
Roller diaphragm of piston pump
Anaesthetic gas scavenging system
Pressure measuring hoses
Connecting hoses for measured
gas return system
Metal parts:
Cylinder head of piston pump
Piston of piston pump
Parts of breathing system
Valves of breathing system
Plastic parts:
Housing of piston pump
Soda line container with screen
T-piece for sidestream gas
measurement
Water trap for measured gas
Water traps for airway
Filters:
Microbial filter
Sensors:
Temperature sensor
(airway temperature)
Temperature sensor
(skin/rectal/oesophagus)
O
sensor (housing)
2
Flow sensor
The following parts cannot be processed and prepared:
(must be replaced if damaged or possibly contaminated)
Filter of pressure measuring hose:
Filter of measured gas return hose:
for sidestream gas measurement:
Water separator:
Connecting hoses:
Filter in T-piece:
Disinfection by
Cleaning with
wiping or immersion automatic cleaners
Wipe
Wipe
externally
Change every 6 months or when resistance increases (e.g. water absorption).
The warning message P-Sensor fault resp. VENT ERR !!! is output.
Replace every 6 months.
Replace, depending on degree of contamination
Replace after each patient
Replace after each patient
Method
Formaldehyde
Steam sterilization
at 121 °C at 134 °C
vapour
Care
Care methods
Tabular summary
24 times
87