Monitor functions
Alarm concept

Heart/lung mode (HLM)

Heart/lung mode (HLM)
A special alarm mode can be activated during operation
for the function »warning« in the »Configuration« menu
for operation with a heart/lung machine (»HLM« mode).
All apnoea alarms are deactivated in this mode and the
gas values are displayed continuously without connection
to the respiratory phase.
SpO
alarms are deactivated in HLM mode. They are
2
automatically reactivated when pulsations are detected
again after ending HLM mode.
In this alarm mode, the parameters are monitored with
the same limits as are set in the ventilation mode from
which »HLM« was selected.
Heart/lung mode can be deactivated either
– via the »Configuration« menu or
– via the softkey »HLM off«.