Preparing for use
Checking readiness for operation with checklist

Vapor

Anaesthetic gas scavenging

Vapor
The illustration shows a Vapor 2000
1 The control dial is engaged at »0«.
2 Filling level OK – check filling level in viewglass.
– Last inspection less than six months ago.
Safety fill:
3 The sealing plug is inserted and secured with
the lever.
Plug adapter:
– The plug adapter lies horizontally and flush all round
on the sealing rings of the plug connection.
Interlock:
4 Plug-in system is locked – locking lever turned
clockwise until it engages.
Anaesthetic gas scavenging
– Is indicator in wall socket green? (Only when using
Dräger systems; note sounds of gas flow in other
cases.)
– Hose connector engaged below breathing system?
– The auxiliary air holes in the tube below the connector
must not be sealed, otherwise the breathing system
will be drained!
Note the separate Instructions for Use for the
anaesthetic gas scavenging system.
22
4
0
1
3
2